CISA released a fact sheet, [Contec CMS8000 Contains a Backdoor], detailing an analysis of three firmware package versions of the Contec CMS8000, a patient monitor used by the U.S. Healthcare and Public Health (HPH) sector. Analysts discovered that an embedded backdoor function with a hard-coded IP address, CWE – 912: Hidden Functionality (CVE-2025-0626), and functionality that enables patient data spillage, CWE – 359: Exposure of Private Personal Information to an Unauthorized Actor (CVE-2025-0683), exists in all versions analyzed.
Please note the Contec CMS8000 may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA’s safety communication, Cybersecurity Vulnerabilities with Certain Patient Monitors from Contec and Epsimed: FDA Safety Communication.
Contec Medical Systems, the company which manufactures this monitor as well as other medical device and healthcare solutions, is headquartered in Qinhuangdao, China. The Contec CMS8000 is used in medical settings across the U.S. and European Union to provide continuous monitoring of a patient’s vital signs—tracking electrocardiogram, heart rate, blood oxygen saturation, non-invasive blood pressure, temperature, and respiration rate. CISA assesses that inclusion of this backdoor in the firmware of the patient monitor can create conditions which may allow remote code execution and device modification with the ability to alter its configuration. This introduces risk to patient safety as a malfunctioning patient monitor could lead to an improper response to patient vital signs.
CISA strongly urges HPH sector organization
[…]